Resting metabolic rate: a comparison between different measurement methods used in male university students
Introduction: Resting metabolic rate (RMR) is a widely used method to determine an individual’s nutritional need. There are different methods of obtaining the RMR values in order to identify these needs. Objective: This study aimed to compare the RMR values using two different indirect calorimetry (IC) devices, bioelectrical impedance analysis (BIA), and two predictive equations in a sample of male university students. Methods: This cross-sectional study analyzed 65 male university students (aged: 23.3 ± 2.8 years, body mass index: 25.78 ± 3.98 kg/m²). RMR was obtained using two IC devices (VO2000® and Fitmate PRO®) and BIA (InBody 570®), and two predictive equations (HarrisBenedict [1918] and FAO/WHO/UNU [2001]). Results: The Bland-Altman (B&A) plot demonstrated considerable differences between the two gold standard methods, VO2000® and Fitmate PRO®, and BIA, with a mean difference of 7.8% and 5.7%, respectively. The B&A plot presented the lowest difference between the two IC devices with a mean difference of 2.1%. The B&A plot also showed a low difference between the two IC devices and predictive equations, with values ranging from −1.3% to −3.6%. In turn, the B&A plot between BIA and the predictive equations comprised approximately −9.2% of the mean difference. The analysis of variance (ANOVA) demonstrated no significant differences between the measurements (p>0.05). Conclusion: Based on the results, InBody 570® underestimates the reference values for RMR in two different IC devices as well as in the predictive equations. Although no significant differences were observed between ANOVA comparisons, the RMR values obtained using BIA should be analyzed with caution. The absence of sophisticated devices for RMR analysis indicates the need to use the FAO/WHO/UNU (2001) equation, which presented a lower difference compared with that of the gold standard measurements in male university students.